Saluki S3302 portable spectrum analyzer is a high frequency handheld instrument. S3302 portable spectrum analyzer has a compact design, very suitable for field measurement. The field strength measurement is indispensable in radiation strength measurement of the tested equipment and can be divided into three modes: PScan, FScan and MScan.
l PScan: Observe the offset, amplitude and field strength of the current point by setting the point frequency.
l FScan: Observe the amplitude and field strength changes within a certain frequency range by setting the starting frequency, step frequency and number of points.
l MScan: Observe the amplitude and field strength of frequency points in the list by recalling the edited or saved list.
1. [PScan]
Main operation procedures of [PScan]:
1) Press [Measure]- [PScan] to enable the point frequency measurement mode.
2) Press [Frequency]- [Frequency] and set the point frequency within the range of [1MHz, 44.1GHz].
3) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.
4) Press [Sweep/Antenna]- [Recall Antenna], and select the antenna factor file.
5) Press [Demod] and set the demodulation type and volume.
The schematic diagram of PScan mode is shown in the figure below (example).
2. FScan
Main operation steps of FScan:
1) Press [Measure]- [FScan] to enable the frequency sweeping mode.
2) Press [Frequency]- [Start Freq] and set the starting frequency of sweeping.
3) Press [Frequency]- [Step Freq] and set the step frequency of sweeping.
4) Press [Frequency]- [Points] and set the number of sweeping points.
5) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.
6) Press [Sweep/Antenna]- [Recall Antenna], and select the antenna factor file.
7) Press [Sweep/Antenna]- [Dwell Time Man], and set the dwell time.
8) Press [Sweep/Antenna]- [Dwell Time Off On] to enable or disable the dwell time.
9) Press [Marker]- [Marker Off On] to enable or disable the marker.
10) Press [Peak] to directly set the marker at the maximum point
The schematic diagram of FScan mode is shown in the figure below (example).
3. MScan
Main operation steps of MScan:
1) Press [Edit List]- [Edit List] and edit the current list.
2) Press [Measure]-[MScan] to enable the MScan mode.
3) Press [Frequency]- [Step Freq] and set the step frequency of sweeping.
4) Press [Frequency]- [Points] and set the number of sweeping points.
5) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.
6) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.
7) Press [Sweep/Antenna]- [Recall Antenna], and select the antenna factor file.
8) Press [Sweep/Antenna]- [Dwell Time Man], and set the dwell time.
9) Press [Sweep/Antenna]- [Dwell Time Off On] to enable or disable the dwell time.
10) Press [Marker]- [Marker Off On] to enable or disable the marker.
11) Press [Peak] to directly set the marker at the maximum point.
The schematic diagram of MScan mode is shown in the figure below (example).