Field Strength Function of S3302 Portable Spectrum Analyzer
Release time :2021.03.14 Reading quantity:453 source:Salukirf, original

Saluki S3302 portable spectrum analyzer is a high frequency handheld instrument. S3302 portable spectrum analyzer has a compact design, very suitable for field measurement. The field strength measurement is indispensable in radiation strength measurement of the tested equipment and can be divided into three modes: PScan, FScan and MScan.

l  PScan: Observe the offset, amplitude and field strength of the current point by setting the point frequency.

l  FScan: Observe the amplitude and field strength changes within a certain frequency range by setting the starting frequency, step frequency and number of points.

l  MScan: Observe the amplitude and field strength of frequency points in the list by recalling the edited or saved list.

1. [PScan]

Main operation procedures of [PScan]:

1) Press [Measure]- [PScan] to enable the point frequency measurement mode.

2) Press [Frequency]- [Frequency] and set the point frequency within the range of  [1MHz, 44.1GHz].

3) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.

4) Press [Sweep/Antenna]- [Recall Antenna], and select the antenna factor file.

5) Press [Demod] and set the demodulation type and volume.

The schematic diagram of PScan mode is shown in the figure below (example).

Fig.1 Schematic Diagram of PScan Mode

2. FScan

Main operation steps of FScan:

1) Press [Measure]- [FScan] to enable the frequency sweeping mode.

2) Press [Frequency]- [Start Freq] and set the starting frequency of sweeping.

3) Press [Frequency]- [Step Freq] and set the step frequency of sweeping.

4) Press [Frequency]- [Points] and set the number of sweeping points.

5) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.

6) Press [Sweep/Antenna]- [Recall Antenna], and select the antenna factor file.

7) Press [Sweep/Antenna]- [Dwell Time  Man], and set the dwell time.

8) Press [Sweep/Antenna]- [Dwell Time Off On] to enable or disable the dwell time.

9) Press [Marker]- [Marker Off On] to enable or disable the marker.

10) Press [Peak] to directly set the marker at the maximum point

The schematic diagram of FScan mode is shown in the figure below (example).

Fig.2 Schematic Diagram of FScan Mode

3. MScan

Main operation steps of MScan:

1) Press [Edit List]- [Edit List] and edit the current list.

2) Press [Measure]-[MScan] to enable the MScan mode.

3) Press [Frequency]- [Step Freq] and set the step frequency of sweeping.

4) Press [Frequency]- [Points] and set the number of sweeping points.

5) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.

6) Press [BW]- [BW] and set the bandwidth as 150Hz, 300Hz, 600Hz, 1.5kHz, 2.4kHz, 6kHz, 9kHz, 15kHz, 30kHz, 50kHz, 120kHz or 150kHz.

7) Press [Sweep/Antenna]- [Recall Antenna], and select the antenna factor file.

8) Press [Sweep/Antenna]- [Dwell Time Man], and set the dwell time.

9) Press [Sweep/Antenna]- [Dwell Time Off On] to enable or disable the dwell time.

10) Press [Marker]- [Marker Off On] to enable or disable the marker.

11) Press [Peak] to directly set the marker at the maximum point.

The schematic diagram of MScan mode is shown in the figure below (example).

Fig.3 Schematic Diagram of MScan Mode